Beta Scintillation Probes
Gamma Scintillation Probes/Data
FLP3B utilizes the same 600 cm²
detector used in the BP17. Mounted in
a wheeled assembly to enable effec-
tive alpha floor monitoring. The
FLM3B is the FLP3B combined with a
Delta-5 ratemeter as seen in picture. A
Separate data sheet is available for
the FLM-3 Series.
LEG-1
Gamma Scintillation Probe Data
Model Application Sensing Element Sensitivity Energy
Element Size (cpm/mR/h) Range
(Dia. x thick)
LEG-1
125
I Measurements NaI(T1) 1” x 0.04” 60 15–200 keV
PG-2
239
Pu, 241 Am, X-ray NaI(T1) 2” x 2 mm 60,000 10–200 keV
SPA-3 High sensitivity gamma NaI(T1) 2” x 2” 1,200,000 60 keV–2 MeV
SPA-6 Medium sensitivity gamma Plastic 50 mm x 60 mm 600,000 40 keV–1.3 MeV
SPA-8 Moderate sensitivity gamma NaI(T1) 1” x 1” 300,000 40 keV–1.3 MeV
SPA-9 Medium sensitivity gamma NaI(T1) 2” x 0.05” 400,000 40 keV–1.3 MeV
This is a gamma scintillation probe for
detection of low-energy gamma and x-
rays, optimized for
125
I, housed in a
lightweight aluminum body. This probe
can be matched with Thermo's portable
survey meters, such as the ASP-2, E-
600, RM-25 or Electra/Delta series.
• Detector type: 1" dia x 0.04" thick NaI
• Window: 75.4 mg/cm² aluminum
window
• Sensitive area: 5.1cm² (0.79 inch
2
)
• Operating voltage: 1,000 V nominal
• Dead time: 8 µs nominal
• Background sensitivity: 60 cpm/mR/h
(
137
Cs)
• Energy Response/Photon: 95%
125
I
• Operating temp: -30 ºC to 60 ºC
(-22 ºF to 140 ºF)
• Connector: CJ-1
• Size: 4.2 dia. x 20.1 cm
(1.65” dia. x 7.9”)
• Weight: 340 g (12 oz)
• Consider as an alternate to the G1LE